Short Term Failure after Residual Intra Mitraclips Regurgitation Treated with Transcatheter Implantation of an Amplatzer Ventricular Septal Defect Device

Luigi Biasco, Marco Moccetti, Francesco Petracca, Elena Pasotti, Daniel Surder, Tiziano Moccetti, Pedrazzini Giovanni

Abstract


We here report the case of a patient with significant recurrent “intraclip” mitral regurgitation following the implant two Mitraclips for a functional mitral incontinence. An AMPLATZER Vascular Plug II deployed between the two clips with acute procedural success with post-procedural trans-esophageal echocardiography (TEE) showing a mild residual mitral regurgitation with trivial trans-valvuar gradients. Few months after the procedure, the patient was newly admitted for acute heart failure due to recurrent MR with a partial detachment of both clips and of the vascular plug forcing a surgical solution for the case.


References


Taramasso M, Zuber M, Gruner C, Gaemperli O, Nietlispach F, Maisano F. First-in-man report of residual "intra-clip" regurgitation between two MitraClips treated by AMPLATZER Vascular Plug II. EuroIntervention. 2014;10:746-52. DOI: 10.4244/EIJY14M12_04

Shewan LG, Coats AJS, Henein M. Requirements for ethical publishing in biomedical journals. International Cardiovascular Forum Journal 2015;2:2 DOI: 10.17987/icfj.v2i1.4




DOI: https://doi.org/10.17987/jatamis.v2i0.364

Refbacks

  • There are currently no refbacks.


Copyright (c) 2017 Luigi Biasco, Marco Moccetti, Francesco Petracca, Elena Pasotti, Daniel Surder, Tiziano Moccetti, Pedrazzini Giovanni

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.